Project and Research Question

Motivation

Automated insulin-delivery research combines software, physiology, noisy sensor data and safety-critical decisions. A controller can improve an average metric while still producing dangerous edge cases. A polished graph can also hide unrealistic rates of change, missing data, excessive interventions or a favourable choice of seed.

IINTS-AF was built to make those problems inspectable. It treats a simulation run as a complete research object: protocol, inputs, model, candidate decision, safety decision, raw outputs, validation and interpretation.

Primary research question

Can an open-source simulation workbench make risky or unrealistic diabetes-algorithm behaviour visible and reproducible before an algorithm is considered for real-world testing?

Hypothesis

An algorithm that reacts to incomplete glucose context can produce questionable actions in stress or failure scenarios. A separate deterministic safety layer, combined with reproducible scenarios and evidence artifacts, should make at least part of this behaviour detectable, blockable and reviewable.

Experimental structure

flowchart LR
    Q["Research question"] --> B["Normal baseline"]
    Q --> S["Physiology or meal stress"]
    Q --> R["Risk or corrupted-input case"]
    B --> C["Candidate and baseline comparison"]
    S --> C
    R --> C
    C --> V["Deterministic validation and safety evidence"]
    V --> E["CSV, metrics, reports and limitations"]

Main contribution

The project contribution is the research workflow around algorithms, not a claim that one controller solves diabetes:

  1. Define a patient, scenario, seed and algorithm.
  2. Run a mechanistic or transparent virtual-patient model.
  3. Keep the candidate algorithm separate from safety authority.
  4. Record candidate and delivered actions.
  5. Validate data and physiological plausibility.
  6. Compare baselines over a locked matrix.
  7. Export raw evidence and human-readable reports.
  8. Allow AI to explain the evidence without changing it.

Why this is different from one notebook

Notebook-only risk IINTS-AF response
Hidden state and manual cells Versioned CLI/SDK workflows and fixed seeds
One attractive run Scenario, patient, algorithm and seed matrices
Black-box output Candidate action, safety action and intervention reason
Metrics without raw evidence CSV traces, manifests, reports and validation files
AI-generated interpretation Deterministic metrics first; LLM explanation second
Unclear data provenance Source manifests, hashes, contracts and MDMP reports
Demo code disconnected from research The demo runs the packaged SDK workflow

Scope of the work

The SDK covers four connected research areas.

1. Simulation

Virtual patients expose latent glucose, insulin action, meal absorption and selected research extensions. A sensor model converts latent glucose into a CGM-like observation.

2. Algorithm and safety experiments

Baselines, experimental controllers and forecasting models can be tested, but the deterministic supervisor remains a separate authority in simulated action paths.

3. Data and AI research

The SDK can prepare diabetes datasets, derive context features, train glucose forecasting models and compare models with both numerical and physiological checks.

4. Evidence and communication

Runs can produce clinical-style reports, AGP-style assets, posters, audit artifacts and a structured evidence bundle for independent review.

Claim ladder

The project uses a claim ladder to prevent results from being overstated.

Level Claim that may be made
Implemented The code path exists and is covered by tests or inspectable output
Verified in software Expected behaviour was observed under specified test conditions
Calibrated in a dataset Parameters or metrics were compared with a documented dataset
Externally validated Performance was tested on an independent population or reference simulator
Clinically validated Requires clinical and regulatory evidence outside the present project

Most IINTS-AF claims are at the first two levels. Some data and realism workflows reach the calibration level. The project does not claim clinical validation.

Success criteria for EUCYS

The EUCYS presentation is successful when a reviewer can answer:

  • What was tested?
  • Which equations and parameters produced the virtual state?
  • What did the candidate propose?
  • Did the safety layer change it?
  • Which output proves that?
  • Can the run be reproduced?
  • Which limitations prevent a clinical claim?