EU Research Software Compliance Dossier

IINTS-AF is research and education software for pre-clinical diabetes technology simulation. It is not a medical device, not a treatment recommendation system, not an insulin dosing tool, and not intended for real-time patient care.

This dossier is a practical governance map for the SDK and the experimental Tauri/Rust desktop shell. It is not legal advice and it is not a conformity assessment. It exists so contributors, reviewers, university partners, and EUCYS-style judges can see how the project handles European research-software expectations.

Regulatory Position

Area Current IINTS-AF position Design consequence
EU AI Act Research-only simulator and local AI analysis workbench; not placed on the market as a clinical AI system. Keep intended purpose narrow; document AI limitations; log runs; preserve human oversight.
Medical Devices Regulation (MDR) Not intended for diagnosis, prevention, monitoring, prediction, prognosis, treatment, or alleviation of disease in a real person. No real dosing use; no treatment advice; no claims of CE/MDR conformity; medical-device boundary review before any clinical use.
GDPR / health data The SDK can process sensitive diabetes data locally when a user imports it. Default local processing, no telemetry, data minimisation, pseudonymisation guidance, no public upload of private datasets.
Cyber Resilience Act / security Open-source research software and downloadable desktop app need security-by-design practices even before formal product obligations mature. Signed releases where possible, dependency scanning, SBOM, narrow Tauri permissions, vulnerability handling.
Research integrity Outputs must be reproducible, reviewable, and not overclaimed. Fixed seeds, manifests, hashes, reports, MDMP certificates, model cards, limitation docs.

Non-Negotiable Boundaries

The SDK and desktop app must always state these boundaries clearly:

  • Research only.
  • Educational simulation only.
  • Not a medical device.
  • Not for insulin dosing, treatment decisions, diagnosis, or real-time patient care.
  • Outputs are hypotheses or research artifacts, not clinical instructions.
  • Local AI may explain or critique runs, but deterministic SDK logs, formulas, and reports remain the source of truth.

EU AI Act Readiness Map

The European Commission describes the AI Act as a risk-based framework and identifies strict obligations for high-risk AI systems, including risk mitigation, dataset quality, logging, technical documentation, deployer information, human oversight, robustness, cybersecurity, and accuracy.

IINTS-AF is not currently positioned as a high-risk clinical AI product. Still, the SDK uses those themes as a research governance checklist:

AI Act theme IINTS-AF control
Intended purpose README, docs, app headers, report disclaimers, and release notes say research-only and not for treatment.
Risk management Safety supervisor, deterministic numeric guards, run validation profiles, realism checks, and local AI red-team review.
Data governance MDMP contracts, dataset registry, dataset license/citation checks, source library, manifest hashes.
Technical documentation Paper technical dossier, physiology reference, architecture overview, formula registry, command reference.
Logging and traceability Run manifests, CSV outputs, audit trails, MDMP certificates, desktop run history.
Transparency Reports explain assumptions, limitations, and whether AI analysis was used.
Human oversight AI suggestions are non-authoritative; deterministic checks and user review are required.
Accuracy/robustness/cybersecurity Pytest, mypy, pip-audit, SBOM, deterministic replay checks, Tauri permission minimisation.

MDR / Medical Software Boundary

The MDR and MDCG software qualification/classification guidance become critical if the project changes intended purpose. The present intended purpose is deliberately outside clinical use.

A formal medical-device review is required before any of these are added:

  • using IINTS-AF to decide insulin or glucagon delivery for a person;
  • connecting output directly to a real pump for treatment;
  • claiming diagnosis, monitoring, prediction, prognosis, treatment, or disease alleviation for an individual patient;
  • claiming clinical performance, safety, CE marking, or MDR conformity;
  • deploying the model into care workflows.

If any of those happen, the project must stop and create a separate regulated-product branch with regulatory expertise, quality management, clinical evaluation, risk management, cybersecurity, post-market surveillance, and conformity-assessment planning.

GDPR / Health Data Controls

Diabetes data is health data. The SDK therefore follows these practical controls:

  • No telemetry by default.
  • Local-first processing by default.
  • Do not commit private CGM, pump, genetics, or clinical files to GitHub.
  • Prefer pseudonymous subject_id values; avoid names, e-mails, dates of birth, MRNs, addresses, and direct identifiers.
  • Keep raw private datasets outside the repo.
  • Public model cards and reports must describe data sources without exposing private records.
  • MDMP certificates can certify data quality, but they do not make private data safe to publish.
  • If real human-subject research is performed, obtain the needed ethics/IRB/DPO approval for the institution and country.

Tauri / Rust Desktop Security Baseline

The Tauri shell should remain a narrow native boundary around the Python SDK:

  • No shell plugin.
  • No broad filesystem plugin.
  • No arbitrary command execution from the frontend.
  • No remote web content in the default app.
  • Fixed Rust command allowlist only.
  • Python bridge accepts fixed subcommands only.
  • Long-running work should run through async Rust commands or background workers.
  • Updates should be signed before automatic deployment is enabled.
  • macOS notarization and Windows signing are release-hardening tasks, not clinical validation.

Release Evidence Checklist

Before a public release, maintainers should check:

  • Full test suite passes.
  • Mypy passes for touched modules.
  • Dependency audit is clean or documented.
  • SBOM is generated.
  • Desktop smoke tests pass on packaged apps.
  • README and app still show research-only/not-medical-device boundaries.
  • Tauri capabilities still do not include shell/filesystem/http/updater permissions unless separately risk-reviewed.
  • Release notes do not claim clinical performance or regulatory approval.
  • Any dataset used in examples has a license/citation and contains no private patient data.

Machine-Checked Controls

The CI governance check covers the core repository controls:

python tools/ci/check_governance.py

That script checks licensing, SBOM shape, dataset registry metadata, manifest hashing, EU research-compliance documentation, privacy boundaries, README disclaimers, and Tauri permission minimisation.

The deeper governance pack is split into:

The JSON control matrix is intentionally machine-readable so CI can verify that each named control still has evidence in the repository.

Runtime Code Guardrails

The SDK also enforces the research boundary in code:

  • src/iints/governance/research_policy.py defines the central research-only notice and post-generation policy scanner.
  • src/iints/analysis/run_quality.py passes local AI run-verifier output through guard_ai_output() before writing reports.
  • src/iints_desktop/local_ai.py passes desktop Ollama answers through the same guard before showing them to users.
  • src/iints_desktop/tauri_bridge.py returns policy-violation labels to the Tauri UI.
  • tools/ci/check_governance.py fails CI if those runtime hooks are removed.

This means a local AI model can still explain research artifacts, but if it tries to provide patient-specific dosing, treatment, diagnosis, or regulatory-approval claims, the SDK replaces the answer with a blocked policy notice.

Official Sources

  • European Commission, AI Act overview: https://digital-strategy.ec.europa.eu/en/policies/regulatory-framework-ai
  • Regulation (EU) 2024/1689, Artificial Intelligence Act: https://eur-lex.europa.eu/eli/reg/2024/1689/oj
  • Regulation (EU) 2017/745, Medical Device Regulation: https://eur-lex.europa.eu/eli/reg/2017/745/oj
  • European Commission MDCG guidance list, including MDCG 2019-11 software qualification/classification: https://health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
  • Regulation (EU) 2016/679, GDPR: https://eur-lex.europa.eu/eli/reg/2016/679/oj
  • European Commission, Cyber Resilience Act overview: https://digital-strategy.ec.europa.eu/en/policies/cyber-resilience-act